Gel chromatography as a method for determining molecular weight. HPLC systems for gel permeation chromatography Basic HPLC system for GPC




Description

Together with the German company Polymer Standards Service (PSS) - one of the leading manufacturers of materials and equipment for gel permeation chromatography (GPC, GPC) or, in other words, size exclusion chromatography (SEC) - we offer complete solutions for the determination of average molecular weights polymers (natural, synthetic, biopolymers), molecular weight distribution and characteristics of polymeric macromolecules in solution. In this method, the separation of the analyte occurs not due to adsorption interactions with the stationary phase, but solely by the value of the hydrodynamic radius of macromolecules.

For the detection of components separated by molecular weight, at least one concentration detector (traditional HPLC refractive and spectrophotometric, evaporative light scattering detector), as well as special detectors for polymer analysis: viscometric, detector by laser light scattering. In combination with concentration detectors, these detectors make it possible to determine the absolute molecular weight, the conformation of macromolecules in solution, the radius of gyration, the hydrodynamic radius, the degree of branching, the constants of the Mark-Kuhn-Houwink equation, and virial coefficients. In the presence of calibration dependences, this system makes it possible to obtain comprehensive information about macromolecular objects and their behavior in solutions in just one analysis (~15 min), while the evaluation of these characteristics by traditional methods takes several days.

To process the measurement results, it is necessary to use special software. We offer flexible, modular HPLC systems for Gel Permeation Chromatography (GPC), including Prominence modules (pumps, column oven, autosamplers, refractive index detector) and specific modules from Polymer Standards Service (PSS), an authority on polymer HPLC analysis. To calculate the results of the analysis, it is possible to use both the Shimadzu GPC Option software integrated into the standard LabSolution LC program, and the use of PSS - WinGPC SW software products that support special detectors.

To work with mobile phases that are aggressive towards traditionally used capillaries and fittings (hexafluoroisopropanol, tetrahydrofuran), HPLC systems can be equipped with a special degasser, pumps and autosampler, the components of which are resistant to these solvents.

Basic systems for GPC

Basic HPLC system for GPC

A basic HPLC system for GPC can be configured with LC-20 Prominence units with one of the concentration detectors (spectrophotometric/diode array SPD-20A/SPD-M20A for UV-absorbing polymers, universal refractive index RID-20A and evaporative light scattering detector ELSD -LTII). This system, in the presence of suitable standards and calibration dependencies, makes it possible to determine the relative molecular weight of polymers, as well as to estimate the hydrodynamic sizes of macromolecules in solution.

Specifications of the main modules
Pump LC-20AD
Pump type Dual Parallel Micro Plunger Mechanism
Plunger chamber capacity 10 µl
Eluent flow rate range 0.0001-10 ml/min
Max pressure 40 MPa
Flow setting accuracy 1% or 0.5 µl (whichever is better)
Ripple 0.1 MPa (for water at 1.0 ml/min and 7 MPa)
Working mode constant flow, constant pressure
The pumps can be equipped with an additional device for automatic flushing of the plunger. The pumps are equipped with a leak sensor. The material of the pump plunger is resistant to aggressive media (sapphire).
Refractometric detector RID-20A
Radiation source Tungsten lamp, operating time 20000 hours
Refractive index range (RIU) 1,00 - 1,75
Temperature control of the optical unit 30 - 60С° with dual optical system temperature control
Operating range of flow rates Ability to work in a wide range of applications (from analytical mode to preparative chromatography) without changing the measuring cell: from 0.0001 to 20 ml/min in analytical mode; up to 150 ml/min in preparative mode
Noise 2.5×10 -9 RIU
Drifting 1×7 -7 RIU/hour
Linearity range 0.01-500×10 -6 in analytical mode
1.0-5000×10 -6 in preparative mode
Flow line switch solenoid valve
Max. operating pressure 2 MPa (20 kgf/cm²)
Cell volume 9 µl
Zero setting optical balance (optical zero);
auto-zero, zero fine-tuning by baseline shift
Column thermostat with forced air convection STO-20A
Controlled temperature range from 10C° above room temperature to 85C°
Temperature control accuracy 0.1C°
The internal volume of the thermostat 220×365×95mm (7.6L)
thermostat capacity 6 columns; in addition to the columns, 2 manual injectors, a gradient mixer, two high-pressure switching valves (6 or 7 ports), a conductometric cell can be installed
Capabilities linear temperature programming; tracking and saving to a file changes in column parameters, the number of analyzes, the amount of the past mobile phase (when installing the optional CMD device)
Performance monitoring solvent leakage sensor; overheat protection system

Light scattering detector

Multi-angle light scattering detector SLD7100 MALLS (PSS)

The SLD7100 MALLS (PSS) multi-angle light scatter detector allows you to measure static light scattering simultaneously at up to seven angles (35, 50, 75, 90, 105, 130, 145°) and determine the absolute values ​​of molecular weights, the true parameters of molecular weight distribution, estimate the size and conformation of macromolecules in solution. This detector eliminates the need for any standards and can also serve as a capacitance instrument (without an HPLC system) without any additional modifications.

Viscometric detector (PSS, Germany)

Viscometric detector DVD1260 (PSS)

The DVD1260 viscometric detector (PSS) when used as part of the LC-20 Prominence HPLC system, allows you to determine average molecular weights and molecular weight distribution parameters, using the universal calibration method, indispensable for macromolecules with complex and globular architecture, as well as the intrinsic viscosity, the constants of the Mark-Kuhn-Houwink equation, the degree of branching, virial coefficients and the conformation of macromolecules in solution, based on certain models already embedded in the software. The unique measuring cell of the detector is a four-arm asymmetric capillary bridge, which, unlike all analogues available on the market, does not contain delay cells (hold-up columns) - a special dilution tank is built into the comparative circuit, which makes it possible to reduce the analysis time by at least half and avoid the appearance of negative systemic peaks. The error of maintaining the temperature in the cell is less than 0.01 °C, which is the first critical factor in viscometric analysis.

Specifications:
Food 110 to 260 V; 50/60 Hz; 100 VA
Differential pressure range (DP) -0.6 kPa - 10.0 kPa
Inlet pressure range (IP) 0-150 kPa
Measuring cell volume 15 µl
Dilution compensation volume (reservoir) 70 ml
Shear rate (1.0 ml/min) < 2700 с -1
Noise level 0.2 Pa, differential pressure signal, 5 °C
analog output 1.0 V / 10 kPa FSD differential pressure
1.0 V / 200 kPa FSD inlet pressure
Total detector volume About 72ml (including tank)
Max. flow rate 1.5 ml/min
Temperature setting accuracy ±0.5 °C
temperature stability Not worse than 0.01 °C
Digital interface RS-232C, USB, Ethernet
Baud rate (baud) 1200 - 115200
Digital inputs Flushing, Zeroing, Injection, Error
Digital outputs Injection, Error
Weight About 4 kg
Dimensions (W, H, D) 160×175×640 mm

Accessories


For work in the GPC mode and construction of calibration dependences, we offer a wide range of speakers for GPC filled with gels (stationary phase) and eluents of a wide variety of chemical nature (polar and non-polar), intended for the analysis of both high molecular weight polymers and oligomers, as well as standard polymer objects.

Gel Permeation Chromatography (GPC, SEC) columns:

  • for any organic eluents: PSS SDV, GRAM, PFG, POLEFIN (up to 200 °C);
  • for aqueous eluents: PSS SUPREMA, NOVEMA, MCX PROTEEMA;
  • columns with monodisperse pore size distribution or mixed type for absolutely linear calibrations;
  • to determine low and high values ​​of MM;
  • ready-made sets of columns to expand the range of determined molecular weights;
  • for synthetic and biopolymers;
  • solutions from micro GPC to preparative systems;
  • columns for quick separations.

Columns can be supplied in any eluent of your choice.

Standards for gel permeation chromatography (GPC, SEC):

  • individual standard samples and ready-made sets of standards;
  • soluble in organic solvents:
    • polystyrene
    • poly(α-methylstyrene)
    • polymethyl methacrylate
    • poly(n-butyl methacrylate)
    • poly(tert-butyl methacrylate)
    • polybutadiene-1,4
    • polyisoprene-1,4
    • polyethylene
    • poly(2-vinylpyridine)
    • polydimethylsiloxane
    • polyethylene terephthalate
    • polyisobutylene
    • polylactide
  • soluble in aqueous systems:
    • dextran
    • pullulan
    • hydroxyethyl starch
    • polyethylene glycols and polyethylene oxides
    • Na-salt of polymethacrylic acid
    • Na-salt of polyacrylic acid
    • Na-salt of poly(p-styrenesulfonic acid)
    • polyvinyl alcohol
    • proteins
  • MALDI standards, validation kits for light scattering detectors (LSD) and viscometry;
  • deuterated polymers;
  • polymers and custom-made standards.

5. Gel chromatography

Gel filtration (synonymous with gel chromatography) is a method for separating a mixture of substances with different molecular weights by filtering through various so-called cellular gels.

The stationary phase in gel chromatography is the solvent located in the pores of the gel, and the mobile phase is the solvent itself, i.e. both the mobile and stationary phases are the same substance or the same mixture of substances. The gel is prepared on the basis of, for example, dextran, polyacrylamide or other natural and synthetic compounds.

Unlike other chromatographic methods that use differences in the chemical properties of the substances to be separated, which manifest themselves during their distribution between the stationary and mobile phases, the separation is based on the sieve effect, which is characteristic of gels with a certain pore radius. The solvent (mobile phase) fills both the external volume between the gel grains and the internal pore volume. The volume of the solvent between the gel grains - V m is called the intermediate, transport or dead volume, and the internal pore volume - V p is considered as an object of the stationary phase. When a sample containing several types of ions or molecules with different sizes is introduced into the column, they tend to penetrate the pores from the mobile phase. Such penetration is due to the entropy distribution, since the concentration of molecules of the substances to be separated in the external solution is higher than in the pore space. But it becomes possible only if the size of the ions or molecules is less than the diameter of the pores.


Figure 5 General view of the calibration curve in gel chromatography:

1 - region of exclusion, where all molecules are larger than m 2 ;

2 - area of ​​penetration or separation, where the sizes of molecules lie in the range from m 1 and m 2;

3 - the area where there is a complete penetration of molecules with sizes less than m 1.

In the process of gel chromatography, large molecules can be separated, which are not adsorbed by the gel, since their sizes exceed the pore sizes, from small ones, which penetrate into the pores, and then can be eluted. Finer separations are also carried out, since the pore sizes can be controlled by changing, for example, the composition of the solvent and, as a consequence, the swelling of the gel. Gel chromatography can be performed in column and thin layer versions.

Used in practice, gels are usually divided into soft, semi-rigid and hard. Soft gels are high-molecular organic compounds with a small number of cross-links. The capacitance factor, equal to the ratio of the volume of the solvent inside the gel to its volume outside the gel, is 3 for them. When swelling, they significantly increase their own volume. These are sephadex or dextran gels, agar gels, starch, etc. They are used to separate mixtures of low molecular weight substances, often in a thin layer version. Chromatography on soft gels is called gel filtration.

Semi-rigid gels are obtained by polymerization. Styrogels, products of the copolymerization of styrene and divinylbenzene with a large number of cross-links, are widely used. The capacity factor of semi-rigid gels is in the range of 0.8 ... 1.2, their volume does not increase very significantly during swelling (1.2 ... 1.8 times). Chromatography on semi-rigid gels is called gel permeation chromatography.

Rigid gels include silica gels and often porous glasses, although they are not gels. Rigid gels have a small capacity factor (0.8...1.1) and a fixed pore size. These materials are used in high pressure gel chromatography.

Solvents for gel chromatography should dissolve all components of the mixture, wet the gel surface and not be adsorbed on it.

The practical application of gel chromatography is mainly associated with the separation of a mixture of high molecular weight compounds, although they are often used for the separation of low molecular weight compounds, since separation by this method is possible at room temperature.

6. High performance liquid chromatography (HPLC)

High performance liquid chromatography is the most efficient method for analyzing complex organic samples. The sample analysis process is divided into 2 stages:

separation of the sample into its constituent components;

· detection and measurement of the content of each component.


The problem of separation is solved using a chromatographic column, which is a tube filled with a sorbent. During the analysis, a liquid (eluent) of a certain composition is fed through a chromatographic column at a constant speed. An accurately measured sample dose is injected into this stream.

The components of the sample introduced into the chromatographic column, due to their different affinity to the sorbent of the column, move along it at different speeds and reach the detector sequentially at different times.

Thus, the chromatographic column is responsible for the selectivity and separation efficiency of the components. By selecting different types of columns, you can control the degree of separation of the analyzed substances. Compounds are identified by their retention time. The quantitative determination of each of the components is calculated based on the magnitude of the analytical signal measured using a detector connected to the output of the chromatographic column.

In the analysis of compounds with low MPCs (biogenic amines, polyaromatic hydrocarbons, hormones, toxins), the sensitivity and selectivity of the method becomes especially important due to the complexity of preparing real samples. The use of a fluorimetric detector makes it possible not only to reduce the limits of detection, but also to selectively isolate the analyzed substances against the background of the matrix and related components of the sample.

The HPLC method is used in sanitary and hygienic research, ecology, medicine, pharmaceuticals, petrochemistry, forensics, quality control and product certification.

The pump "Python" of the syringe type is used as the eluent supply unit, which has the following features:

· absence of pressure pulsations when supplying the solvent;

a wide range of volumetric flow rates;

large volume of the pump chamber;

Extensibility (the ability to combine several blocks to create a gradient system).

Different types of detectors can be used in the chromatographic system, for example, "Fluorat-02-2M" (spectral selection is carried out by filters) or "Fluorat-02 Panorama" (spectral selection is carried out by monochromators).

7. Application

Liquid chromatography is the most important physical and chemical research method in chemistry, biology, biochemistry, medicine, and biotechnology. It is used to analyze, separate, purify and isolate amino acids, peptides, proteins, enzymes, viruses, nucleotides, nucleic acids, carbohydrates, lipids, hormones, etc.; study of metabolic processes in living organisms of drugs; diagnostics in medicine; analysis of products of chemical and petrochemical synthesis, intermediates, dyes, fuels, lubricants, oils, wastewater; study of isotherms of sorption from solution, kinetics and selectivity of chemical. processes.

In the chemistry of macromolecular compounds and in the production of polymers, liquid chromatography is used to analyze the quality of monomers, study the molecular weight distribution and distribution by type of functionality of oligomers and polymers, which is necessary for product control. Liquid chromatography is also used in perfumery, the food industry, for the analysis of environmental pollution, and in forensics.


Conclusion

The beginning of the 20th century was marked by the discovery of the chromatographic method of analysis, which enriched and united various fields of science, without which the scientific progress of the 21st century is unthinkable. The introduction of chromatographic methods, and primarily liquid chromatography, into medicine has made it possible to solve many vital problems: the study of the degree of purity and stability of drugs, the preparative isolation of individual hormonal drugs (for example, insulin, interferon), the quantitative determination of neurotransmitters in biological objects: adrenaline, norepinephrine. The presence of these substances in a living organism is associated with the ability to memorize, learn, acquire any skills. HPLC identification of steroids, amino acids, amines, and other compounds proved to be extremely important in the diagnosis of certain hereditary diseases: myocardial infarction, diabetes, and various diseases of the nervous system. One of the urgent tasks of clinical medicine for express diagnostics is the so-called profile analysis of the components of a biological object, carried out by liquid chromatography, which makes it possible not to identify each peak, but to compare chromatogram profiles to conclude about the norm or pathology. The processing of a huge amount of information is carried out only with the use of a computer (the method is called the "pattern recognition method").


Bibliography

1. V. P. Vasiliev, Analytical Chemistry, in 2 books. Book. 2 Physical and chemical methods of analysis: Proc. for stud. universities studying chemical engineering. specialist. - 4th ed., stereotype. – M.: Bustard, 2004 – 384 p.

2. Moskvin L.N., Tsaritsyna L.G. Separation and concentration methods in analytical chemistry. - L.: Chemistry, 1991. - 256 p.

3. http://bibliofond.ru/view.aspx?id=43468

4. http://ru.wikipedia.org/wiki/Paper_chromatography

5. http://referats.qip.ru/referats/preview/93743/6

6. http://www.curemed.ru/medarticle/articles/12186.htm

7. http://www.lumex.ru/method.php?id=16

8. http://www.xumuk.ru/encyklopedia/1544.html

9. http://www.pereplet.ru/obrazovanie/stsoros/1110.html

transcript

1 INSTITUTION OF THE RUSSIAN ACADEMY OF SCIENCES INSTITUTE OF ELEMENTO-ORGANIC COMPOUNDS im. A.N. NESMEYANOV. SCIENTIFIC AND EDUCATIONAL CENTER FOR PHYSICS AND CHEMISTRY OF POLYMERS MOSCOW

2 Table of contents. BASES OF CHROMATOGRAPHY OF POLYMERS. Driving forces and modes of polymer chromatography. Chromatographic peak characteristics. The concept of theoretical plates..3 Fundamentals of the size-exclusion (gel-penetrating) chromatography method. CARRYING OUT PRACTICAL WORK ON THE ANALYSIS OF MWD OF THE POLYMER BY THE METHOD OF GEL PERMEATION CHROMATOGRAPHY 3. REFERENCES. BASICS OF POLYMER CHROMATOGRAPHY. Driving forces and modes of polymer chromatography. Chromatography is a method of separating substances by distributing between two phases, one of which is mobile and the other is immobile. The role of the mobile phase in liquid chromatography is played by a liquid (eluent) moving in channels between particles along a column filled with a porous material (see Fig.). Fig. Movement of a macromolecule in a chromatographic column: d k - the size of the channels between the particles of the stationary phase; dn - pore size; R is the size of the macromolecule; t s - time spent by the macromolecule in the pore, t m ​​- in the mobile phase. The stationary phase is the pores of the sorbent filled with liquid. The average velocity of movement of this phase along the axis of the column is equal to zero. The analyte moves along the axis of the column, moving along with the mobile phase and occasionally stopping when it enters the stationary phase. This process is illustrated in Fig., which schematically shows the jump-like motion of a macromolecule with size R through channels with size d corresponding to the particle size. Molecules stop in slit-like pores, the size of which corresponds in order of magnitude to the size of macromolecules. The time between successive stops can be written as:

3 t t s + t m + t k, () where t s is the residence time of the molecule in the stationary phase, t m ​​d is the time spent by the molecule in the mobile phase (D - D is the transverse diffusion coefficient, t k is the time of transition from the mobile phase to the stationary phase and vice versa). Usually in the processes of high performance liquid chromatography (Hgh Performance Lqud Chromatography in the English literature) in its analytical version, this time t k is much less than the first two and can be omitted in the formula (). If the number of stops during the movement along the column is sufficiently large, then the total time of the macromolecule movement along the column is sufficiently large as compared to the characteristic time of equilibrium establishment. In this case, to determine the probability of finding a macromolecule in a unit volume of the stationary phase with respect to the mobile phase (or the distribution coefficient K d equal to the ratio of concentrations in these phases), the methods of equilibrium thermodynamics can be used. Namely, the distribution coefficient will be determined by the free energy of the transition of the macromolecule from the mobile phase to the stationary phase: T S H G RT Kd exp exp () RT For a chain consisting of N segments, K exp(N µ), (3) d where µ is the change in the chemical potential segment. The distribution coefficient in chromatography is a fundamental concept and is defined as follows: VR V K d (4) Vt V t is the elution volume of the substances leaving together with the solvent front. From (3), one can immediately see that, depending on the sign of G, the macromolecules behave differently when they enter the pore (see figure): Fig.. if G>, then K d tends to with increasing length of the macromolecule the volume of elution also decreases). This corresponds to size exclusion chromatography. At G< K d экспоненциально растет с ростом ММ и это соответствует адсорбционному режиму хроматографии. Таким образом, оба режима хроматографии могут рассматриваться в рамках единого механизма и, более того, плавно меняя энергию взаимодействия сегмента с поверхностью сорбента за счет состава растворителя или температуры, можно обратимо переходить от одного режима к другому. Экспериментально это было впервые показано в работе Тенникова и др. . Точка (для данной пары полимер - сорбент - это состав растворителя и температура), соответствующая равенству G, при которой происходит компенсация энтропийных потерь и энергетического выигрыша при каждом соударении сегмента макромолекулы со стенкой поры называется критической точкой адсорбции или критическими условиями хроматографии. Как видим, в этих условиях не происходит деления по ММ и это обстоятельство является предпосылкой для использования режима критической хроматографии для исследования разных типов молекулярной неоднородности полимеров, таких как число функциональных групп на концах цепи, состав блоксополимеров, топология 3

4 (presence of branched or cyclic macromolecules). This chromatographic method is relatively new and some of the most interesting results of its application can be found, for example, in [,3,4]. Chromatography mode corresponding to condition G< широко применяется для разделения низкомолекулярных соединений и называется, в зависимости от химической природы функциональных групп на поверхности сорбента, адсорбционной, нормальнофазной, обращеннофазной, ионпарной и т.д. хроматографией. Для полимеров его применение ограничено областью слабых взаимодействий вблизи критических условий и областью олигомерных макромолекул, т.к. с ростом длины цепи мы переходим к практически необратимой адсорбции макромолекулы на колонке. Наиболее важным для полимеров является режим эсклюзионной хроматографии или, как его еще называют, гельпроникающей хроматографии. Этот режим более подробно будет рассмотрен в следующем разделе, а сейчас мы перейдем к описанию некоторых важнейших хроматографических характеристик... Характеристики хроматографического пика. Концепция теоретических тарелок. После прохождения через хроматографическую колонку узкой зоны какого-либо монодисперсного вещества, на выходе мы получаем расширенную зону в виде пика приблизительно гауссова по форме (в случае хорошо упакованной колонки и правильно выбранной скорости хроматографии). Причины расширения пика лежат в различных диффузионных процессах, сопровождающих движение молекул вдоль колонки (см. например, соотношение ()). Наиболее важные характеристики пика - объем элюирования или V R или объем удерживания (относится к центру пика) и дисперсия пика, т.е. второй центральный момент (см.рис.3): σ h V V dv R. (5) Справедливы следующие соотношения между величинами, показанными на рис.3: σ, 43W W b. (6) 4 Рис. 3. Модель гауссова пика. Параметры уширения пика. Часто все эти величины выражаются в единицах времени, тогда говорят о времени удерживания и т.д., однако, в этом случае скорость потока элюента должна быть строго фиксирована. Существует простая феноменологическая теория описания относительного вклада расширения зоны в хроматографическое разделение. Это - теория тарелок. Хроматографическая колонка мысленно делится на ряд последовательных зон, в каждой из которых достигается полное равновесие между растворенным веществом в подвижной и неподвижной фазе. Физическую основу этого подхода составляет скачкообразное движение, описанное в начале первого раздела, и число теоретических тарелок в колонке связано с числом остановок при попадании в неподвижную фазу за время движения данного вещества по колонке. Чем больше это число, тем больше число теоретических тарелок и тем выше эффективность колонки. Число теоретических тарелок определяется следующим образом: 4

5 VR N σ V 5.54 W R V 6 W R b. (7) Since this value changes with the elution volume, it is correct to use the unretained substance exiting at K d..3 to characterize the efficiency of the column. Fundamentals of the size-exclusion (gel-penetrating) chromatography method. Size exclusion chromatography (Sze Excluson Chromatography, SEC) or gel permeation chromatography (GPC, Gel Permeaton Chromatography, GPC) is implemented when the behavior of macromolecules in pores is determined by the entropy component of free energy, and the energy component is small compared to it. In this case, the distribution coefficient will depend exponentially on the ratio of macromolecule size and pore size. The scaling theory predicts the following regularities for the case of pores commensurate with the macromolecule size R K d Aexp D α, (8) 4/3 to depending on the adopted pore model (slit, capillary, strip) and the chain model (ideal or imperfect). Thus, the behavior of macromolecules under the conditions of size exclusion chromatography is determined by the chain size. The size of a macromolecule is determined by its chemical structure, the number of links in the chain (or molecular weight), topology (for example, the size of a branched macromolecule or macrocycle decreases compared to a linear macromolecule of the same chemical structure). In addition, the size of flexible macromolecules depends to a certain extent on the solvent used due to the excluded volume effect. However, the GPC method has become widely used in laboratory practice as a method of separation by molecular weights, determination of average molecular weights and molecular weight distributions (MWDs). The development of the method began in the mid-1950s, when the first wide-pore organic sorbents for high performance gel permeation chromatography were created. As can be seen from relations (8), the method is not absolute for determining molecular weights, but requires an appropriate calibration against standard (preferably narrowly dispersed) samples with known MW, relating the retention volume (or time) to MW. Figure 4 illustrates the calibration curves for polystyrene in terms of lg V R on Waters semi-rigid organic sorbents (crostyragel) with different pore sizes. For the analysis of any polymer by molecular weight, it is necessary to select a column with an appropriate pore size or a series of columns with different pores, or use a column with a mixture of sorbents with different pores (the Lnear column in the given example). Of course, in order to use the GPC method for the analysis of MWD, it is necessary to provide conditions for the implementation of the exclusion mechanism of separation, which is not complicated by the effects of the interaction of both middle and terminal links of the chain. We are talking about adsorption interaction from a nonpolar solvent or reversed-phase interaction of nonpolar chain fragments during chromatography of hydrophilic polymers in an aqueous medium. In addition, water-soluble polymers containing ionized groups are capable of strong electrostatic interactions and require especially careful selection of chromatography conditions. The selection of conditions includes the selection of a sorbent and solvent (eluent) suitable for a particular analysis in terms of chemical structure. 5

6 Recommendations can be found in the manuals of chromatographic equipment manufacturers, as well as in reference books and monographs (see, for example, ), 6 V R, ml Pic. 4. Calibration curves for µstyragel columns. The figure shows the corporate labeling of the columns with a value that characterizes the size of the sorbent pores, which is equal to the length of the extended polystyrene chain excluded from the pores for steric reasons. The chromatographic column is the heart of the liquid chromatograph. The chromatograph also includes a number of necessary additional devices:) an eluent supply system (pump) that provides a stable flow,) a sample injection system without stopping the flow (injector or autosampler), 3) a detector - a device that provides the formation of a signal proportional to the concentration of a substance at the column outlet (detectors are of various types, the most popular in gel permeation chromatography are refractometric and spectrophotometric detectors), and 4) data acquisition and processing systems based on a personal computer. In modern chromatographs, the operation of all parts of the chromatograph is often also controlled by means of a control program integrated with the data processing system. The polymer chromatogram obtained under size exclusion chromatography F(V) is a reflection of its molecular weight distribution function W(). By virtue of the law of conservation of matter: F V dv W d ). The real chromatogram is the result of the separation of the sample by MW when moving along the column and the simultaneous mixing of polymer homologues due to the blurring of the zones. Therefore, the function F(W) in relation (9) should be understood as a chromatogram corrected for PU. This function is a solution to the Fredholm integral equation of the first kind. There are quite a lot of ways to correct for PU. See, for example, . However, in modern high-performance chromatographic systems, in most cases, the contribution of PU to the chromatogram is small compared to MWD and can be neglected. The most important procedure is the calibration of the chromatograph according to the molecular weight of the polymer under study. If there are corresponding narrowly dispersed standards with different MM, the elution volumes (V R or Ve) are determined for them and a calibration dependence similar to that shown in Fig. 4 is built. Typically, the calibration relation is sought in the form (): n lg C V e () Polynomials of the first or third degree are most often used. Polynomials of odd degrees (3, 5, 7) most accurately describe the characteristic shape of the calibration curves with upper and lower MM limits. Sets of narrowly dispersed standards exist for such polymers as polystyrene, polyisoprene, polymethylmethacrylate,

7 polyethylene oxide, dextrans and some others. You can also use the method of universal calibration, first introduced into practice by Benoit and co-workers. The method is based on the fact that the hydrodynamic volume of macromolecules is proportional to the product of the intrinsic viscosity and the molecular weight of the polymer and can be used as a function of the elution volume as a universal parameter for different polymers. Then we build a universal gauge relation (), () lg η n BV e, () using a set of some standards and the well-known Mark-Kuhn-Houwink relation (3): η K a. (3) To pass from a relation of the form () to a calibration dependence () for the polymer under study, it is sufficient to use the corresponding Mark-Kuhn-Houwink relation, after which we obtain (4): lg n B V e + a lg K. (4) As a result, From the data of gel permeation chromatography, one can find the average molecular weights of various degrees of averaging, which, by definition, are the following values: () n - number average MM, W () d W d w z W d W d W d W d - weight average MM, - z-mean MM. The MM ratios of different degrees of averaging characterize the statistical width of the MMD. The most commonly used ratio is w / n, which is called the polydispersity index. 4. CARRYING OUT PRACTICAL WORK ON THE ANALYSIS OF MWD OF A POLYMER BY METHOD OF GEL PENETRATION CHROMATOGRAPHY Purpose of the work: To get acquainted with the operation of a liquid chromatograph, the method of conducting a chromatographic experiment, the method of calibrating a chromatograph according to narrowly dispersed polymer standards and calculating average molecular weights. Equipment:) Liquid chromatograph, consisting of a pump, an injector, a column thermostat, a column with a polymeric sorbent and a data processing system based on a personal computer.) A set of narrowly dispersed standards with different MM (polystyrene or polyethylene oxide). 3) Test sample with unknown molecular weights. Operation procedure:) Preparation of a solution of a mixture of standards. 7

8) Obtaining a chromatogram of the standards and determining their retention volumes (V e). 3) Construction of the calibration dependence in the form (). 4) Preparation of a solution of the investigated polymer. 5) Obtaining a chromatogram of the investigated polymer. 6) Calculation of the average MM of the sample. Figure 5 shows a typical example of a polymer sample chromatogram prepared for calculating the average MW, namely, a baseline is drawn that defines the beginning and end of the chromatogram, and then the chromatogram is divided into equal parts along the time axis, the so-called slices. n w z A, A A A, A A. 5. For each slice, its area A is determined and the molecular weight corresponding to its middle is calculated from the calibration dependence. The average molecular weights are then calculated: 8

9 3. LITERATURE. M. B. Tennikov, P. P. Nefedov, M. A. Lazareva, S. Ya. Comm., A, 977, v.9, N.3, with S.G.Entelis, V.V.Evreinov, A.I.Kuzaev, Reactive oligomers, M: Chemistry, T.M.Zimina, E.E. Kever, E.Yu. Melenevskaya, V.N. Zgonnik, B.G. Belenkiy, On the experimental verification of the concept of chromatographic "invisibility" in the critical chromatography of block copolymers, Vysokomolek. comm., A, 99, vol. 33, N6, with I.V. Comm., A, 997, v.39, N6, with A.M. Skvortsov, A.A. Gorbunov, Scaling theory of chromatography of linear and ring macromolecules, Vysokomolek. comm., A, vol. 8, N8, with B. G. Belenkiy, L. Z. Vilenchik, Chromatography of polymers, M: Chemistry, W. W. Yau, J. J. Krkland, D. D. Bly, orn Sze-Excluson Lqud Chromatography, New York: John Wley & Sons, E.L. Styskin, L.B. Itsikson, E.B. Braudo. Practical High Performance Liquid Chromatography. Moscow Ch Wu, Ed.Column Handbook for Sze Excluson Chromatography, N-Y: Academc Press..Z.Grubsc, R.Rempp, H.Benor, J. Polym. Sc., B, 967, v.5, p


ESTABLISHMENT OF THE RUSSIAN ACADEMY OF SCIENCES INSTITUTE OF ELEMENTORGANIC COMPOUNDS im. A.N. NESMEYANOV. RESEARCH AND EDUCATIONAL CENTER FOR PHYSICS AND CHEMISTRY OF POLYMERS Blagodatskikh I.V. LIQUID CHROMATOGRAPHY OF POLYMERS

1 Macromolecular compounds (Lysenko EA) Lecture 7. Fractionation of macromolecules 2 1. The concept of fractionation. 2. Preparative fractionation. 3. Method of turbidimetric titration. 4. Gel-penetrating

Laboratory work 7b Chromatographic determination of the composition of the gas phase of soils. Chromatography (from the Greek chroma, genitive chromatos color, paint) is a physicochemical method of separation and analysis

8. Questions 1. Define chromatography. 2. What features of chromatography make it possible to achieve a better separation of substances with similar properties compared to other separation methods. 3. List

Moscow Institute of Physics and Technology (State University) Department of Molecular Physics Physical research methods Lecture Gas chromatography Theory and principles Dolgoprudny, November

04.07 Moscow Institute of Physics and Technology Department of Molecular and Biological Physics Physical research methods Lecture 8 Chromatography Dolgoprudny, April 6, 07 Plan. History of occurrence

Moscow Institute of Physics and Technology (State University)) Department of Molecular Physics Physical Methods of Research Lecture 0 Gas Chromatography Dolgoprudny, November 5, 0g. Plan. Story

Analytical chemistry 4th semester, Lecture 17. Module 3. Chromatography and other methods of analysis. Chromatography. Principle and classification of methods. 1. The principle of chromatographic separation. Stationary and mobile

Discovery of chromatography (1903) MIKHAIL SEMENOVICH TsVET (1872-1919) Main stages in the development of chromatography 1903 Discovery of chromatography (Tsvet M.S.) 1938 Thin-layer or planar chromatography (Izmailov

Moscow Institute of Physics and Technology (State University) Department of Molecular and Biological Physics Physical research methods Lecture 7 Gas and liquid chromatography. Practical

CHAPTER 7 GAS-LIQUID CHROMATOGRAPHY As a method of analysis, chromatography was proposed by the Russian botanist MS Tsvet to solve the particular problem of determining the components of chlorophyll. The method turned out to be universal.

Moscow Institute of Physics and Technology Department of Molecular and Biological Physics Physical Research Methods Lecture 9 Gas Chromatography Experimental Technique and Methods Dolgoprudny, April 3

Topic 5. Fundamentals of rheology. Viscosity of polymer solutions. Theoretical part. Viscous liquids and solutions of macromolecular substances (NMWs) are divided into Newtonian and non-Newtonian according to the nature of the flow. Newtonian

Benefits of Agilent AdvanceBio SEC SEC columns for biopharmaceutical analysis Comparing columns from different vendors to improve data quality Technical Overview

Moscow Institute of Physics and Technology (State University) Department of Molecular and Biological Physics Physical research methods Lecture 9 Liquid chromatography Methods and technology

Journal of Analytical Chemistry, 5, volume 6, 7, p. 73-78 UDC 543.544 Simulation of gas chromatography for a given dependence of Henry's constant on temperature. 5g Prudkovsky A.G. Institute of Geochemistry and Analytical

Agilent AdvanceBio SEC Size Exclusion Chromatography Columns for Aggregation Analysis: Instrument Compatibility Technical Overview Introduction Agilent AdvanceBio SEC columns are a new family

MULTI DETECTOR GEL PERMEATION CHROMATOGRAPHY for POLYMER ANALYSIS K. Svirsky, Agilent Technologies, [email protected] Gel Permeation Chromatography is the only chromatographic technique that

ANNOTATION of the work program of the discipline "Introduction to chromatographic methods of analysis" in the direction of preparation 04.03.01 Chemistry on the profile of training "Analytical chemistry" 1. The objectives of mastering the discipline

46. ​​CHROMATOGRAPHIC SEPARATION METHODS Chromatographic separation methods are multi-stage separation methods in which the sample components are distributed between two phases, stationary and mobile. motionless

MINISTRY OF EDUCATION AND SCIENCE OF RUSSIA NATIONAL RESEARCH TOMSK STATE UNIVERSITY FACULTY OF CHEMISTRY Annotated work program of the discipline Chromatographic methods of analysis Area of ​​study

Scientific and technological company SINTEKO METHOD OF QUANTITATIVE CHEMICAL ANALYSIS OF COFFEE AND TEA FOR CAFFEINE CONTENT BY METHOD OF LIQUID CHROMATOGRAPHY. DZERZHINSK 1997 1 This document is distributed

Lecture 7 (9.05.05) TRANSFER PROCESSES IN GAS

FEDERAL AGENCY FOR EDUCATION State Educational Institution of Higher Professional Education “Ural State University. A.M. Gorky" IONTS "Ecology and nature management"

Macromolecular compounds (Lysenko E.A.) Lecture 5 (-Temperature). - temperature and ideality of the solution. - temperature and phase equilibria. 3. - temperature and sizes of macromolecular coils. .. Influence

Lecture 6 Chromatographic methods of analysis Lecture plan 1. Concepts and terms of chromatography. 2. Classification of chromatographic methods of analysis. Chromatographic equipment. 3. Types of chromatography: gas,

Theory of real matter. Science presents a large number of theories or laws of a real gas. The most famous van der Waals real gas law, which increases the accuracy of the behavior description

BELARUSIAN STATE UNIVERSITY FACULTY OF CHEMISTRY DEPARTMENT OF ANALYTICAL CHEMISTRY

Lecture 7. SURFACE PHENOMENA 1. Surface tension 1.1. surface energy. Until now, we have not taken into account the existence of an interface between different media*. However, its presence can be very

Viscoelasticity of polymeric liquids. Basic properties of polymeric liquids. Highly entangled polymer fluids include polymer melts, concentrated solutions, and semi-diluted

Moscow Institute of Physics and Technology (State University)) Department of Molecular Physics Physical research methods Lecture 9 Chromatography. Introduction Dolgoprudny, 9 October 0g. Plan.

ANALYTICAL CHEMISTRY UDC 543.544 ADSORPTION CHROMATOGRAPHY IN BIOGAS ANALYSIS 1999 M.V. Nikolaev Research Institute of Chemistry, UNN N.I. Lobachevsky L.P. Prokhorova Nizhny Novgorod aeration station A technique has been developed

MODERN PREPARATIVE FLASH CHROMATOGRAPHY Part 2* A.Abolin, Ph.D., "GalaChem" [email protected] P.-F. Ikar, Interchim (France) We continue to publish materials on modern methods of preparative

Quick Guide to Selecting Columns and Standards for Gel Permeation Chromatography SELECTION GUIDE Introduction Gel Permeation Chromatography (GPC) is a technique for evaluating molecular weight distribution

MINISTRY OF HEALTH OF THE RUSSIAN FEDERATION GENERAL PHARMACOPEIAL AUTHORIZATION Chromatography OFS. SP XI, issue 1 Chromatography is a method for separating mixtures of substances, based

Agilent Software for Gel Permeation Chromatography One-Stop Solution for Fast and Easy Polymer Analysis Key Features Introduction Agilent Technologies

2.2.29. HIGH PERFORMANCE LIQUID CHROMATOGRAPHY High performance liquid chromatography (HPLC) is a separation technique based on the differential distribution of substances between two immiscible

Yaroslavl State Pedagogical University. K. D. Ushinsky Department of General Physics Laboratory of Molecular Physics Laboratory work 5 Study of statistical regularities on the Galton board

Lecture 3. FREE SURFACE ENERGY OF THE PHASE DIVISION Surface forces. Surface Tension Consider a system containing liquid and vapor in equilibrium with it. Density distribution in the system

2 Methods of analysis: 1. Chemical methods. Chemical equilibrium and its use in analysis. Acid-base balance. The strength of acids and bases, the patterns of their change. Hammet function. calculation

Lecture 7 Branched chain reactions. Critical phenomena in branched chain reactions. E.-K. pp. 38-383, 389-39. A simple expression for the rate of formation of radicals: d r f(p) g(p) (1)

Lecture 6 Lukyanov I.V. Transport phenomena in gases. Contents: 1. The length of the free path of molecules. 2. Distribution of molecules over their mean free paths. 3. Diffusion. 4. Viscosity of the gas (internal friction).

Federal State Budgetary Institution of Science "Kirov Research Institute of Hematology and Blood Transfusion of the Federal Medical and Biological Agency" 3.3.2. Medical immunobiological

1. Explanatory note 1.1. Requirements for students The student must have the following initial competencies: basic provisions of mathematical and natural sciences; master the skills of independent

1 LECTURE 10 Two systems in diffusion contact. chemical potential. Phase equilibrium condition. Heat of transition. Clausius-Clapeyron formula. Two systems in diffusion contact Equilibrium state

1. A list of competencies indicating the stages (levels) of their formation. PC-1: the ability to use knowledge of the theoretical, methodological, procedural and organizational foundations of forensic science, forensics

Topic. Physico-chemistry of surface phenomena. Adsorption. Surface phenomena manifest themselves in heterogeneous systems, i.e. systems where there is an interface between components. Surface phenomena

TOMSK STATE UNIVERSITY Faculty of Physics STUDYING THE VISCOSITY COEFFICIENTS OF A LIQUID BY THE STOKES METHOD Guidelines for performing laboratory work Tomsk 2014 Reviewed and approved

Highly elastic polymer meshes. polymer meshes. Polymer networks consist of long polymer chains linked together and thus forming a giant three-dimensional macromolecule. All polymer

Gas chromatography 1 Substance requirements 1. Volatility 2. Thermal stability (substance must evaporate without decomposition) 3. Inertness Gas chromatograph layout 1 2 3 4 5 1. Carrier gas cylinder

The curriculum is based on the OSVO 1-31 05 01 2013 educational standard and the HEI curriculum G 31 153/ac. 2013 COMPILER: V.A.Vinarsky, Associate Professor, Candidate of Chemical Sciences, Associate Professor RECOMMENDED

MINISTRY OF HEALTH OF THE RUSSIAN FEDERATION GENERAL PHARMACOPIES AUTHORIZATION Chromatography on paper OFS.1.2.1.2.0002.15 Instead of Art. SP XI, issue 1 The chromatographic process occurring on the filter sheet

Ministry of Education and Science of the Russian Federation Federal Agency for Education State Educational Institution of Higher Professional Education "UFA STATE OIL

AGILENT POLYMER STANDARDS FOR GPC/SEC Contents POLYMER STANDARDS FOR GPC...3 InfinityLab EasiVial...5 InfinityLab EasiCal...8 Polystyrene Standards...9 Standards

GROUP P A C O M P A N I B I O C H I M M A K Z A C R B I O 1 1 9 8 9 9, Russia, Moscow, Lenin Tel./Fax (0 9 5 ) 939-59-67, tel. 939- I N S T R U C T I A ​​on the use of the Analytical Kit MOSCOW

Theory of Ion Chromatography: A Universal Approach to Describing Peak Parameters 1998 A.G. Prudkovskii, A.M. Dolgonosov V. I. Vernadsky Institute of Geochemistry and Analytical Chemistry, Russian Academy of Sciences, 117975

Moscow Institute of Physics and Technology (State University) Department of Molecular and Biological Physics Physical research methods Lecture 8 Detectors in chromatography Liquid chromatography

MINISTRY OF HEALTH OF THE RUSSIAN FEDERATION GENERAL PHARMACOPEIAL AUTHORIZATION Electrophoresis OFS.1.2.1.0021.15 Instead of Art. SP XI, issue 1 Electrophoresis method of analysis based on the ability of charged particles,

1 Macromolecular compounds (Lysenko E.A.) Lecture 10. Thermomechanical analysis of amorphous polymers. 2 1. Basic concepts of mechanical analysis of physical bodies. 2. Thermomechanical curves of amorphous polymers

5 PHYSICAL EQUILIBRIUM IN SOLUTIONS 5 Partial molar values ​​of mixture components Consideration of the thermodynamic properties of a mixture of ideal gases leads to the relationship Ф = Σ Ф, (5) n where Ф is any extensive

6. Moscow Institute of Physics and Technology (State University) Department of Molecular Physics Physical research methods Lecture Gas chromatography. Technical implementation Liquid chromatography

Macromolecular compounds (Lysenko EA) Lecture 4. Phase equilibria in polymer solutions. Dissolution kinetics. Concentration modes. Equation of state of polymer solution. . Phase equilibria

Laboratory work. Determination of the content of C 8 arenes in the gasoline fraction Knowledge of the hydrocarbon (HC) composition of oils and condensates at the molecular level is of great importance both for petrochemistry

Ideal polymer chain. Ideal polymer chain. An ideal chain is a model chain in which the so-called bulk interactions are neglected, i.e. interactions of links removed along the chain.

Laboratory work 1.17 STUDYING THE LAW OF NORMAL DISTRIBUTION OF RANDOM VARIABLES M.V. Kozintseva Purpose of the work: to study the distribution of random variables on a mechanical model (Galton board). Exercise:

BELARUSIAN STATE UNIVERSITY APPROVED Dean of the Faculty of Chemistry D.V. Sviridov 2011 Registration UD- /r SOLUTIONS OF POLYMERS Curriculum in the specialty 1-31 05 01 Chemistry (by directions)

Gel chromatography as a method for determining molecular weight

Gel Permeation Chromatography is a type of column fractionation method in which the separation is carried out according to the molecular sieve principle. This principle was already known in the early 1950s, but it was only after Porat and Flodin rediscovered and widely used this method that it was recognized and widely used in scientific research. From that moment until 1964, more than 300 papers were published on this new fractionation method.

Gel filtration or size exclusion chromatography(sieve, gel permeation, gel filtration chromatography) - a type of chromatography, during which the molecules of substances are separated in size due to their different ability to penetrate into the pores of the stationary phase. In this case, the largest molecules (of larger molecular mass) that are able to penetrate into the minimum number of pores of the stationary phase are the first to leave the column. Substances with small molecular sizes, which freely penetrate into the pores, come out last. In contrast to adsorption chromatography, in gel filtration, the stationary phase remains chemically inert and does not interact with the substances to be separated. The stationary phase is the pores of the sorbent filled with liquid. The average velocity of movement of this phase along the axis of the column is equal to zero. The analyte moves along the axis of the column, moving along with the mobile phase and occasionally stopping when it enters the stationary phase. Molecules stop in slit-like pores, the size of which corresponds in order of magnitude to the size of macromolecules.

In size exclusion chromatography, molecules that are large in solution either do not penetrate at all, or penetrate only part of the pores of the sorbent (gel) and are washed out of the column earlier than small molecules. The ratio of the effective sizes of macromolecules and pores of the sorbent determines the distribution coefficient Kd, which determines the retention volume of the component VR in the column:

The effective size of a macromolecule in size exclusion chromatography is its hydrodynamic radius R, which together with the polymer molecular weight M determines the intrinsic viscosity of the polymer. The universal calibration dependence of V R on the product / equation (2) was first obtained experimentally by G. Benois, it has the form (Fig. 1):

where A and B are constants. Equation (2) is equally valid for linear and branched polymers, block and graft copolymers, and oligomers.

Rice. one.

molecular size exclusion chromatography

In the region from V 0 to V T (column volume available for the solvent and molecules below a certain size, corresponding to M min), the working dependence is linear (quasi-linear) in nature. Corresponding volumes V 0 and V T mol. masses represent the exclusion limits - M max (large molecules, do not penetrate into the pores of the sorbent) and M min, (molecules are small, completely penetrate into the pores of the sorbent). Sorbents with pores of the same size are theoretically capable of separating macromolecules within the limits commercial sorbents are characterized. To separate macromolecules in a large range of M, sorbents with a bimodal and trimodal pore size distribution are needed, providing a linear mol. mass calibration dependence in the range М = 10 2.5 - 10 6.5 . Maximum selectivity is achieved by increasing the volume of the sorbent's pore space for bimodal and trimodal sorbents, in addition, by optimal pore size distribution. It is important that when separating a mixture of macromolecules, their largest and smallest M should be within the limits of M MIN - M MAX characteristic of a given sorbent.

Mechanism of size exclusion chromatography. Size Exclusion Chromatography (SEC) or Gel Permeation Chromatography (GPC, Gel Permeation Chromatography, GPC) is implemented when the behavior of macromolecules in pores is determined by the entropy component of free energy, and the energy component is small compared to it. In this case, the distribution coefficient will depend exponentially on the ratio of macromolecule size and pore size. Macromolecules in p-re are statistical. ensemble (statistical tangle). Their distribution between the porous sorbent and the solution is controlled by the change in the Gibbs energy during the transition of the macromolecule from the solution to the pores: where is the change in the enthalpy of the macromolecule due to the interaction. its segments with the surface of the sorbent (gel matrix); - decrease in entropy during the transition of the macromolecule from the solution to the pores; T - abs. t-ra. The separation of macromolecules occurs in the exclusion mode, when a K d , which depends on the ratio of the sizes of macromolecules and pores, is less than 1. To suppress the phenomena of ion exclusion and ion exchange sorption, which are undesirable for size exclusion chromatography, the surface of the sorbents is modified (to give it a neutral charge at pH > 4) , increase the ionic strength of the solvent, weakening the Coulomb interactions, add organic solvents, thereby shifting the pK of the polyelectrolyte or the isoelectric point of polyampholytes. On the other hand, ion exchange sorption and ion exclusion can be used to separate neutral macromolecules, polyanions, and polycations of the same size. Since the dissociation of polyelectrolytes increases with dilution of their solutions, during size exclusion chromatography, macromolecules at the edges of the chromatographic column, where their concentration is low, dissociate and move along the column not according to the laws of size exclusion chromatography, but according to the laws of ion exchange sorption and ion exclusion, depending on the charge of the sorbent surface and macromolecules, which leads to a distortion of the shape of the curve of the dependence of V and M (Fig. 2), and also makes it possible to diagnose the presence of one or another process.

Rice. 2. Size exclusion chromatography of neutral macromolecules (a) and polyelectrolytes: ion exclusion (b), ion exchange sorption (c)

Effects similar to ion-exchange sorption, but only to a lesser extent, can be observed during hydrophobic interactions of macromolecular segments with a sorbent surface modified by hydrophobic radicals or during electrostatic interactions of surface silanol hydroxy groups with functional groups of polar macromolecules. All these effects must be suppressed by size exclusion chromatography.

To analyze any polymer by molecular weight, it is necessary to select a column with a suitable pore size or a series of columns with different pores, or use a column with a mixture of sorbents with different pores (the Linear column in the given example). Of course, in order to use the GPC method for the analysis of MWD, it is necessary to provide conditions for the implementation of the exclusion mechanism of separation, which is not complicated by the effects of the interaction of both middle and terminal links of the chain. We are talking about adsorption interaction from a nonpolar solvent or reversed-phase interaction of nonpolar chain fragments during chromatography of hydrophilic polymers in an aqueous medium. In addition, water-soluble polymers containing ionized groups are capable of strong electrostatic interactions and require especially careful selection of chromatography conditions. The selection of conditions includes the selection of a sorbent and solvent (eluent) suitable for a particular analysis in terms of chemical structure.

Size exclusion chromatography technique. To separate macromolecules in size exclusion chromatography, two types of columns are used: those operating in narrow = 10 2) and wide (= 10 4 - 10 5) ranges. Columns with a wide M range have a wide sorbent pore size distribution (bimodal, trimodal). This distribution is selected in such a way that, for a given degree of linearity of the calibration mol.-mass dependence and mass range, the greatest degree of selectivity is provided. Size exclusion chromatography is carried out using a chromatograph, the detector is a spectrophotometer or a flow refractometer with a sensitivity limit of 5 x 10 -8 units. refraction, which corresponds to a polymer concentration of 5-10 -5%. Typically, the instrument operates at room temperature, however, polyolefin size exclusion chromatography requires an elevated temperature, which increases separation selectivity, column efficiency, and analysis speed due to a decrease in the viscosity of the mobile phase. Modern chromatographs are equipped with an automatic device for preparation (polymer dissolution, solution filtration) and sample injection, a computer for interpreting the results of MMP analysis. The use of a combination of a refractive index detector and a photometer makes it possible to determine the MWD and branching indices without calibrating the chromatograph according to polymer standards. During gel filtration of proteins, it is necessary to take measures to prevent their adsorption on the sorbent and prevent their denaturation. Unlike size exclusion chromatography of synthetic polymers and oligomers, which is mainly used for analytical purposes, protein gel filtration is one of the most important methods for their isolation and purification.

Size exclusion chromatography uses macroporous inorganic or polymeric sorbents. For size exclusion chromatography of polar polymers, inorganic sorbents (silica gels and macroporous glasses) are modified with organosilicon radicals, and for size exclusion chromatography of hydrophilic polymers, with hydrophilic groups. Among the polymeric sorbents, styrene-divinyl-benzene sorbents are the most common (for size exclusion chromatography of high polymers and oligomers). For gel filtration of biopolymers, primarily proteins, hydrophilic polymer sorbents (sephadexes - dextrans with cross-links, as well as polyacrylamide gels) or macroporous silica gels modified with polysaccharides are used.

Size exclusion chromatography is effectively used in the development of new polymers, technological processes for their production, production control and standardization of polymers. Size exclusion chromatography is used to analyze the MWD of polymers, research, isolation and purification of polymers, including biopolymers.

The physical basis of this method is very simple and clear. The investigated polymer solution flows through a column filled with a porous sorbent. The separation of component mixtures is based on the distribution of the substance between the mobile (flowing solvent) and stationary (solvent in the pores of the sorbent) phases, i.e., on the different ability of polymer macromolecules to penetrate into the pores of gel granules, hence the name of the method.

The surface of the sorbent granules is covered with many channels, depressions and other irregularities, conditionally called pores, the total volume of which is V„. The volume inaccessible to the solvent is called the dead volume. Let a solution flow past such a surface, the size of which is commensurate with the size of the pores or less than them. Some of these molecules penetrate into the pores if their concentration in the moving phase is greater than in the pores. When the solute zone leaves this area of ​​the sorbent, the concentration of molecules inside the gel pores becomes greater than outside, and the molecules again diffuse into the flow of the mobile phase. If the size of the molecules is larger than the size of the pores, then such a molecule passes by the gel granule without lingering, i.e., it is excluded (exclusion) from the pore space. Thus, larger macromolecules flow through the column faster. This means that different polydisperse sample molecules will exit the column at different times for different retention volumes. VR

VR= V0 +kvV>

Where Vo- volume of the mobile phase (current solvent); kv- pore volume distribution coefficient: for large macromolecules completely excluded from pores, kv = 0; for solvent molecules kv= 1),

Values VR depend mainly on temperature, the nature of the solvent and the concentration of the solution.

The behavior of a macromolecule in solution can be easily described in detail if its Gibbs energy is determined AG. If a macromolecule enters a pore, its entropy decreases. In the presence of interaction of segments of the macromolecule with the walls of the pore, the enthalpy changes: with attraction, the enthalpy decreases, and vice versa. Therefore, in the absence of adsorption AG > 0, with strong adsorption of macromolecules on the pore walls AG < 0. Accordingly, in the first case, size exclusion chromatography (size distribution) takes place, and in the second - adsorption; conditions at AG=0 called critical. Because in the region AG > 0, macromolecules are separated by size, analysis by molecular weights of linear polymers is possible. If the polymer is branched, the separation process becomes more complicated and depends on the type and number of branches, and in the case of copolymers, also on the composition and blockiness of the chain.

Gels of hydrophobic materials, such as polystyrene cross-linked with divinyl-benzene, have received the greatest use as sorbents: In such gels, the effects of adsorption of analyzed samples are almost completely absent. Recently, macroporous glasses have been widely used, which have a number of advantages compared to a polymeric sorbent (particle rigidity, pore size variation, chemical stability) and disadvantages (increased sorption of polymers on them).

The most commonly used solvents are tetra-hydrofuran (THF), chloroform, toluene, cyclohexane and mixtures thereof. Preference is given to THF, which, unlike toluene, does not form micelles or aggregates with polymer macromolecules and is transparent in the UV region of the spectrum. In addition, the efficiency of method 11IX using THF is maximum at rather low temperatures (35–45°C). However, during long-term storage, THF oxidizes with the formation of explosive peroxide compounds; therefore, it must be pre-purified. Using THF as a solvent, all rubber grades as well as thermoplastic elastomers can be analyzed. When carrying out the analysis of nitrile butadiene rubber, it is advisable to use a mixture of solvents, one of which has an affinity for the non-polar rubber unit, and the other for the polar one. If a refractive index detector is used, a necessary requirement for the solvent is the difference between the refractive indices of the solvent and the polymer.

For the first time, an instrument for gel chromatography poly analysis Merov was released by "Waters" in 1964, later five years after Method discoveries. Today, liquid chromatographs for analysis Molecular weight distribution (MWD) polymers are produced in all industrialized countries, in Russia chromatographs of the KhZh series are known. Among the latest modifications of foreign devices is a gel chromatograph manufactured by "Waters Chem. Div." with a viscometer to determine the molecular weight, MWD, as well as the degree of orientation of macromolecules. The carousel design of the device allows you to simultaneously test 16 samples.

The block diagram of the chromatograph includes: О Degasser block - serves to remove gases from the solvent and helps to maintain the same amount of solvent for a long time.

О Dispenser block - allows you to enter a sample of a given volume in time and work in automatic mode,

О In modern liquid chromatographs, the conversion of the chromatogram into the MWD of the polymer, including instrument calibration by molecular weight and correction for instrument broadening, is carried out using a computer. This allows us to calculate the differential and integral MWD and the average values ​​of the molecular weight using the accepted programs. Special microprocessors control the operation of the device blocks according to a given program.

An example of recording the conditions of an experiment conducted by the method of gel permeation chromatography. The installation consists of the following main elements; pump model 6000A, sampler U 6K and differential refractometer R 401. The unit also includes 3 separating columns ^ each 300 mm long and with an internal diameter of 8 mm. The columns are filled with SDV-Gel 5 which has a pore diameter of 103, 104 and 105 A (Polymer-Standard-Service, PSS, Mainz). The test temperature is 22° C. and the flow rate is 1.0 ml/min. Tetrahydrofuran is used as a solvent, the injection volume is 100 µl at a sample concentration of 6-10 g/l. Universal calibration is performed on polystyrene with a molecular weight of 104-106 g/mol.

GPC allows you to study subtle changes in the chemical structure of polymers and determine the total MWD, and therefore is widely used in polymer chemistry. In the industrial production of elastomers, the GPC method can be used for operational quality control of commercially available products and appropriate adjustment of the technological process, as well as in the development and improvement of obtaining elastomers with desired properties. Gel chromatographs can be included in automated process control systems with sampling for analysis directly from the reactor. The duration of the analysis, including sample preparation, is 20-30 minutes.